KLEINWACHTER CPM374 CHARGED PLATE MONITOR 
with integrated Electro-Fieldmeter 
SYSTEM PROF. DR. ING. HANS KLEINW?CHTER 
 
The CHARGE PLATE Monitor 374 measures the voltage potential on its integrated Plate-Electrode according the Fieldmill-Influence-Principal and is suitable to determine the Static Decay Time from a selected higher voltage threshold to a lower voltage threshold level. 
  
CPM374 
Forschungs-. Entwlcklungs-. Produktlons-u. Vertrlebsges.m.b.H
 
The CHARGE PLATE Monitor CPM374 consists of an integrated timer, a high-voltage unit and an electro-fieldmill-sensor and achieves therefore the high impedance of 10
15 W to measure electrostatic voltage potentials and static decay times of air ionizer equipment. 
 
The unit meets the requirements of the Standard IEC 61340-5-1 and the American Standard EOS/ESD S.3.1.  
 
For the vertical measurement use the included three Distance pieces. Screw them on the backside of the unit. 
 
Important: The high voltage unit is protected by a high ohmic current limiting resistor and therefore there is no danger when touching the charged plate. 
 
The CPM374 is battery powered and due to its compact design, it is very suitable for the use in industrial areas. It can also be operated at a power supply (230/115 VAC). The CPM374 is menu-driven and will be controlled by means of 2 switches for simple operation. All selected parameters will be shown in the 2 rows LCD-display for a better orientation of the operator.  
 
The enclosed software enables the remote control operation of the CPM374 and data processing by means of a PC. 
 
2.1  Switching on  
The CHARGE PLATE Monitor CPM374 is equipped with a NiCd-battery and ready for operation - use the enclosed power supply only to charge the battery. The display background illumination is not operating when battery powered. The battery operation features an automatic switch-off function, if the switches were not touched for 
approximay 10 minutes. when the unit is run on the power supply (230/115 VAC) the background illumination is switched on while the automatic switch-off function is disabled. 
 
Press button B to switch on the CPM374. 
  
1.  Description 
2.  Operation 
2.2  Decay Time and Offset Voltage 
 
The software version will be shown shortly before the measurement mode and the command to initiate the measurement appear on the display: 
 
DECAY TIME: AUTO 
PRESS START! 
 
The initially setting is saved in the EEPROM and will be shown after each switch-on. 
 
If the auto-mode is selected, as indicated in the example above, then the time will be measured for the positive voltage decay first and then followed automatically for the negative. This mode is recommended for checking air ionizer equipment, as the Startbutton has to be pushed on time only to perform a complete check of ionizers including the determination of the Offset Voltage. 
 
Offset-Voltage is the voltage potential, which remains after a certain time on the plate electrode. It is an indication for the balance of positive and negative ions. If the Offset- Voltage is positive, then an excess of positive ions emanate from the ionizer equipment - the other way round for negative. Please consider the specification of the ionizer for a qualified evaluation. 
 
The Start-button has to be pressed in order to charge the plate-electrode by means of the integrated high-voltage unit. The timer setting and the applied voltage potential are shown on the display: 
 
TIMER: 1000-100V 
U = XXXXV 
 
The timer starts as soon as the upper threshold value (1000 volt in this case) is passed: 
 
TIMER=XX.Xs 
U = XXXXV 
 
and will stop the decay time when reaching the lower value (100 volt). The voltage setting for the timer could be changed in the set-up function. 
 
The offset-voltage will be measured after a certain lapse time at the end of the static decay time measurement: 
OFFSET=XXXs 
U = XXXXV 
 
The measurement series is now finished and all relevant parameters can be obtained from the display, e.g. to qualify an air ionizer equipment. The offset-voltage UOFF and the belonging lapse time will be finally displayed in the first row for 10 seconds or until a new measurement was initiated by pressing button A. 
 
 The decay time appears in the second row as POS DECAY or as NEG DECAY or both  alternating every second, if the auto-mode was pre-selected: 
 
UOFF=XXXV T=XXXs 
POS.DECAY =XX.Xs 
 
2.3  Saving measurement values  
The measured values can be saved in the internal memory, therefore the request will follow after each measurement accordingly:  
SAVE VALUES? 
A<YES NO>B 
 
press the button A to start the saving. In case that the memory is full (99 entries), the display will indicate for approximay 2 seconds: 
 
MEMORY FULL! 
 
All files have to be deleted in order to perform the saving of the last measurement: The following request will appear on the screen:  
DELETE FILES? 
A<YES NO>B 
 
In case, that the full memory should not be deleted, please press B and the unit will be ready to take the next measurement - the last measurement results will disappear and cannot be saved. 
 
Important: It is recommended to transfer all files to your PC after each measurement interval by means of the enclosed software. Thus will allow to delete files of 
the CPM374 at the next measurement interval without loosing important data. 
 
Press A to delete the full memory of the CPM374 and the proposal for the file number of the last measurement will appear: 
 
FILE NO: XX 
A<OK CHANGE>B 
 
Press A to confirm this file number or change with each press of B until you reach the right file number. Finally, press A to confirm the right file number. 
In order to reach faster a certain file number press B, then the ten digit can be changed by pressing A while the single digit can be changed with B. when you have reached the right file number, wait for approximay 3 seconds until the display change back as stated above and confirm the setting by pressing A. 
 
Press A and B simultaneously to start the set-up function. The CPM374 will ask for performing the set-up function:  
SET-UP ? 
A<YES NO>B  
 
3.1  SET-UP  
If the set-up function was started, the last Static Decay and Offset setting will appear (in this case Mode = Auto), which can be confirmed again by pressing A:  
MODE = AUTO 
A<OK CHANGE>B 
 
If B will be pressed, the following options are shown one after the other: 
MODE = POS 
MODE = NEG 
MODE = VOLTMETER 
MODE = AUTO 
 
The decay time for only positive respectively negative potentials with the following offset voltage are measured in the MODE = POS and MODE = NEG. In the MODE = 
VOLTMETER, the CPM374 will measure permanently the voltage potential on the plate electrode - Use a measurement lead connected to the plate electrode to operate the 
CPM374 as an Electrometer (High Impedance Voltmeter), e.g. monitoring the change of voltage potential on human bodies, caused by movements. 
 
If you have selected the decay time measurement (MODE = POS, MODE = NEG or MODE = AUTO), the starting point for the voltage potential, from where on the decay time 
should be measured, will be asked first:  
START: U=1000V 
A<OK CHANGE>B 
 
Press A to confirm the proposed starting point. If B will be pressed, the following options are shown one after the other:  
500V - 550V - ... - 900V - 950V - 1000V - 500V 
 
and the selected value has to be confirmed by pressing A . 
 
 
3.  SET-UP Function  
The same procedure will follow selecting the stop point, that means at which potential the decay time should be stopped:  
STOP: U=100V 
A<OK CHANGE>B 
 
Press A to confirm the proposed stop point. If B will be pressed, the following options are shown one after the other:  
0V - 50V - 100V - ... -400V - 450V - 500V - 0V 
 
and the selected value has to be confirmed by pressing A .  
Finally, the CPM 374 will ask for the lapse time to determine the offset voltage:  
OFFSET: XXXs 
A<OK CHANGE>B 
  
Press A to confirm the proposed lapse time. If B will be pressed, the following options are shown one after the other:  
10s - 20s - ... - 50s - 1min - 2min- ... - 10min - 10s 
 
and the selected value has to be confirmed by pressing A  . 
The chosen parameters for Start, Stop and Offset will be shown for approximay 3 sec, 
e.g.: 
TIMER: 1000-100V 
OFFSET: XXXs 
 
and finally the unit will recheck if all parameters are chosen correctly:  
TIMER+OFFSET OK? 
A<YES NO>B 
 
The selected parameters will be confirmed by pressing A and could be changed again with B. 
When A was pressed, the unit will ask for clearing the memory, as the stored data files are based on the previous setting:  
CLEAR MEMORY? 
A<YES NO>B 
 
 
If B was pressed, the previous setting will remain the same, the memory was not deleted and the new parameter change is ignored, the following message will appear: 
 PARAMETER 
NOT CHANGED ! 
 
If A was pressed, then following message will advice that the parameter change was taken over and the memory is cleared accordingly, the following message will appear for approximay 2 sec:  
PARAMETER 
CHANGED ! 
 
The display will show now the new decay time setting and the request to press start! 
 
3.2  Calibration  
If NO>B was chosen for the set-up function, the unit will ask for the Electro-Fieldmeter calibration:  
EFM CALIBRATION? 
A<YES NO>B 
 
Press B if a calibration is not required. Press A to initiate the automatic offset calibration, the following message will appear for approximay 1 sec:  
PLEASE WAIT 
CALIBRATION 
 
Important: The plate electrode must be free of foreign voltage connection during calibration. 
 
The display will show the new decay time setting and the request to press start! 
 
If the electrofield-meter is defected, the following message will appear:  
EFM DEFECT! 
REPAIR IS NEEDED 
 
If the power of the NiCd Battery is less than 6,7 volts, the unit will provide the following message alternating every second with the actual first row: 
 
LOW BATTERY!  
 
4.  Other messages on the display 
The battery has to be recharged, but the ongoing or the upcoming measurement could be finished first.  
 
Important:  
Please use the enclosed battery charger only for charging the batteries. The batteries will also be charged when the CPM374 is switched off. The max. charging time is 14 hours, but the unit could be longer connected to the battery charger as it consists of a charging control device. It is recommended to use the CPM374 until LOW BATTERY! appears on screen. This will maintain the max. battery capacitance over a longer period. 
 
If the battery voltage will become less than 6,1 volts, the following message will appear on screen for approximay 2 seconds:  
AUTO OFF 
LOW BATTERY! 
 
and the unit will switch off automatically to prevent a total discharge of the battery. A total discharge can cause damage or destroy the battery! 
Connect the unit by means of the provided cable to a COM socket of a PC. Insert the floppy disk and select the right drive (e.g. A:). Type ?START“ and press ENTER, the software will be installed on your PC. 
 
Requirements: 
min. 486 processor, 3.5“ floppy disk drive, 2 MB RAM, DOS or Windows. the program is self explaining. 
We provide 12 month warranty from date of receiving the unit, if the unit was handled properly. The damage of the NiCd-battery caused by improper handling as well as 
mechanical damage on the unit are excluded from warranty. Warranty is not provided, if the housing was opened. 
 
5.  PC-Operating 
6.  Warranty 
The unit has to be grounded sufficiently for accurate measurements of the electrical field and its polarity. Therefore the unit has to be connected to ground potential by means of its grounding socket. 
 
Housing :  Aluminum-housing 152mm x 152mm x 152mm (L x W x H) 
Weight :  1,5 kg 
Display :  Alphanumeric Display, 2 rows with 16 digits each, 
display area 100 mm x 24 mm 
Measurement Range :  Voltage potentials up to ± 1200 V, accuracy ± 2,5% 
PC-Interface :  Serial port, 9 pol Sub D – socket 
Battery :  7,2 V / 850 mAh 
last approximay 4 hours for continuous operation when 
battery was fully charged, charging time max. 14 hours 
Battery Charger :  12 V / 500 mA 
Load Current:  max. 350mA, max. 150 mA for battery operating 
Operating Temperature:  0 ... 40°C 
Relative Humidity:  0 ... 60% 
 
·  CHARGE PLATE Monitor CPM 374 
·  Power transformer prim: 230VAC sec: 12VDC 500mA 
·  Teflon insulated measurement lead, 1m 
·  Crocodile clip 
·  Grounding cable 
·  Serial PC-port cable 
·  3 Spacers for horizontal decay time measurement 
·  Operating Instructions 
·  Certificate of factory calibration 
 
  
7.  Grounding 
8.  Specification 
9.  Parts of Delivery 
 
The measurement unit is not approved for the use in explosive areas! 
The unit is not approved for the use in power utilities! 
This unit is not suitable to measure alternating fields with a frequency higher than 1 Hz. 
Sparking discharge to the electro-fieldmill-sensor has to be avoided. 
 
The Teflon-supports of the plate electrode must be maintained clean, therefore do not touch with fingers. All Teflon spacers and the teflonisolated wire to the plate have to be cleaned carefully with Acetone or similar solvents in case that the voltage on the charged plate drops in a rate of greater than 100V / minute under normal circumstances (no external lead connection to the plate electrode and no presence of air ionizers). The interval for factory calibration is normally 1 year. 
 
 
Please note the following advice when checking air ionizers: 
The presence of charged objects as well as personnel could cause failure on the measurement results caused by influence of voltage. The unit should be therefore always 
grounded to avoid this phenomena from the housing of the unit itself. 
 
 
10.  Safety Advice 
11.  Maintenance 
12.  Important Notice 
 
KONFORMIT?TSERKL?RUNG 
 
Die Firma 
 
Kleinw?chter GmbH 
Krummattstr. 9 
79688 Hausen i.W. 
 
erkl?rt in alleiniger Verantwortung, dass das Produkt 
 
Charge Plate Monitor CPM 374 
Messger?te zur Messung des elektrischen Gleichspannungsfeldes 
 
auf die sich diese Erkl?rung bezieht, die Forderung 
 
der EWG-Richtlinie über die elektromagnetische Vertr?glichkeit 
Richtlinie des Rates vom 3.5.1989 (89/336) Stand: Juli 1993 
 
erfüllt, insbesondere der Normen 
 
EN 61010 ?Safety“ 
EN 55011 Gruppe 1, Klasse B 
EN 50082-2/92 
 
für ISM-Ger?te. 
 
Die Firma Kleinw?chter GmbH h?lt folgende technischen Dokumentationen zur Einsicht bereit 
 
·  vorschriftsm??ige Bedienungsanleitung 
·  Baupl?ne 
·  technische Dokumentationen 
 
Hausen, den 09.01.2002 
 
 
Jürgen Brunner 
techn. Gesch?ftsführer 
 
 
KLEINWACHTER
Forschungs-. Entwlcklungs-. Produktlons-u. Vertrlebsges.m.b.H


中國電子行業(yè)儀器儀表供應(yīng)商——上海堅(jiān)融實(shí)業(yè)有限公司      :www.0798tv.com  
凌志堅(jiān) 產(chǎn)品     :        
:jet1718  地址:上海市閔行區(qū)閔北工業(yè)園區(qū)紀(jì)王鎮(zhèn)紀(jì)西周涇 9 號/68 號 

上一篇:德國Kleinwachter TOM-600靜電電阻測試儀手冊說明書

下一篇:DESCO50671/50672離子風(fēng)機(jī)操作說明書手冊

產(chǎn)品分類

聯(lián)系我們

在線客服:

亚洲欧美唯美另类丝袜-欧美日韩一区伊人-国产一级做a爱片久久片-2016亚洲天堂网 | 久久国产精品99久久久久久口爆-激情五月天色婷婷-麻豆在线免费视频观看-黄色片一级视频网站 | 天堂蜜桃一区二区特点-亚洲婷婷在线一区-日韩午夜激情毛片-婷婷六月天丁香综合 | 日韩在线观看免费网-免费中文字幕一区二区-中文日韩一区二区三区四区五区-日韩激情欧美在线,激情网 | 中文字幕成人av在线-97人妻精品一区二区三区动漫-亚洲中文字幕人妻系列-国产免费久久久9999 | 日韩美女免费视频啪-17c视频在线观看一区二区-久热这里只有精品9-18岁日韩内射颜射午夜久久成人 色婷婷观看免费视频-国产精品自产拍在线-国产成人福利在线视频-亚洲国产精品国自产拍张津瑜 | 懂色av中文一区二区三区-日本japanese少妇好高清-欧美91在线免费观看-久久二人精彩视频免费观看 麻豆天美传媒av国产在线观看-av在线播放一区二区三区-国产模特一区二区三区-日韩在线免费观看a级视频 | 久久久亚洲熟妇熟女一区二区-伊人婷婷涩六月丁香七月-欧美中文字幕精品人妻影-国产又大又圆又黄的视频 | 久久草青青在线视频-91色婷婷综合久久久中文-国产一级黄在线观看-又硬又粗又长又大我好想要 | 97人妻精品一区二区三区-一区二区三四区免费观看熟女视频-蜜乳av一区二区三-日韩精品卡通动漫网站 | 91麻豆精品亚洲永久-亚洲a资源在线观看-乱子伦一区二区三区视频在线观看麻-超碰97资源在线免费观看 | 日韩精品在在线视频-欧美精品成人a在线观看麻豆-熟妇人妻中文字幕老熟妇-久久美女福利视 | 中文字幕日韩一区二区-日韩免费av中文字幕一区二区-99热99这里只有精品6首页-久久人妻天天操天天干 | 91免费福利网尤物在线播放-五月婷婷中文开心字幕-日韩av懂色av-成人动作片中文字幕 | 国产99re在线播放-麻豆亚洲a√熟女国产-成人生活片国产一级-午夜区二区三区不卡 | 国产一区二区三区22-欧美日韩熟女制服丝袜-久久久久久av资源-日韩不卡一区二区视频 | 日本国产在线播放一区二区不卡-风流老熟女视频一区二区-伊人久久久最新地址-蜜桃久久国产一区二区 | 国产在线看片免费视频97-日韩欧美一区av-国产中文字幕在线观看免费-天天干夜夜操激情黄色 | 777色婷婷av一区二区三-81精品国产乱码久久久久久-91久久久久免费版-国产成人av一区二区三区不 | 亚洲成人 自拍偷拍-91熟女性高潮视频-日韩黄片视频在线免费看-超碰免费人人妻 | 日韩一区二区在线视频观看-国产18禁久久精品-99久久久99久久91熟女-国产乱子伦一区二区三区视频播放 | 中文字幕熟女乱-亚洲三级成人在线观看-日韩 欧美 制服 丝袜 中文-亚洲av一区在线播放 | 91福利国产一区-日韩av一区的-蜜臀av精品一区二区三区-久久精品久久久久久国产,免费 | 中文字幕人妻诱惑在线播放-久久嫩草精品久久久精品内容-中文字幕国产有码av-久久久精品国产亚洲乱码 | 一区二区三区四区五区伊人-av中文字幕免费看-国产欧美日韩免费成人-婷婷激情五月手机版 | 精品人妻少妇一区二区三区不卡-国产亚洲欧美日韩男男网站-99久久夜色精品国产-亚洲精品久久久久久久久久久中文字幕 | 国产精品,精品视频-日本极品少妇的粉嫩sex-国内又粗又爽又黄视频在线观看-亚洲精品日韩国产精品 国产一区二区打炮视频-1024一区二区三区人妻-国产精品97色视频-人妻中出视频在线观看 | 国产69精品久久久黑-久久狠狠爱亚洲综合-日韩精品亚洲激情-国产乱码久久久久久国产 | 成人av亚洲av日本av-97精品国产97久久久久久春色-久久久精品国产亚洲成人满-国产一区二区三区视频在线 | 91大神康先生在线-久久久亚洲av明日花绮罗-69精品人妻一区二区三区蜜桃-国产一区二区三区又黄又爽 | 久久色亚洲av-久久久久久久久91精品一区二区-中文字幕精品丝袜人妻-欧美日韩专区一区二区 | 久久精品国产91精品亚洲高清-精品人妻久久久蜜桃-日本人妻中文字幕精品专区-欧美久久久久久久高潮 | 久久九九精品视频-午夜精品一区二区三区四区五区-激情久久亚洲精品-91色视频最新地址 | 99热,这里只有精品免费-国产精品久久久久7777婷婷-成人av在线一区二区-男人操女人逼穴网站 | 久色视频在线观看免费-99久久久国产成人精品-久久视频黑少妇-久久综合乱子伦国产免费88 | 久久久人妻91久久久久久-亚洲综合色一区二区三区蜜臀-日一区二区三区在线视频-日韩福利视频一区二区三区 | 成人性生交大片免费看午夜-狠狠综合久久av一区二区…-国产91丝袜在线18-亚洲伊人狠狠另类综合 | 国产精品99久久久久久www-麻豆郭童童在线播放-亚洲人妻高清一区二区三区-97国产免费电影网 | 婷婷婷婷久久久久-久久久精品日本人妻-久久久久久久精品婷婷-99国产精品视频在线 | 中文字母人妻一区二区三区69-蜜臀av久久久久久-69色在线视频播放-天天干天天日天天操天天色天天射 | 精品人妻一区二区免费看-久久嫩草精品在线-日熟女一区二区三区-97超碰在线观看日 |